Exact 1-port VNA error correction and a GUM uncertainty budget, computed in your browser.
| Short |S11| (linear) | 0.9947 |
| Short angle (deg) | 186.1 |
| Open |S11| (linear) | 0.9679 |
| Open angle (deg) | 4.13 |
| Load |S11| (linear) | 0.03 |
| Load angle (deg) | 20 |
| DUT raw |S11| (linear) | 0.2221 |
| DUT raw angle (deg) | 45.65 |
| Source match (return loss, dB) | 26 |
1-port OSL correction: DUT 0.2221∠45.65deg (raw) -> 0.20001∠45.01deg corrected, RL 13.98 dB, VSWR 1.500, U=+/-0.01498 (k=2). Exact 3-term Moebius solve, residual 2.22e-16 -> RESID_BELOW_1E10. Standards: ideal (Open=+1, Short=-1, Load=0). A QA/documentation companion to a calibrated VNA + traceable standards, not a replacement.
| Corrected DUT |Gamma| | 0.20001 |
| Corrected phase (deg) | 45.01 |
| Return loss (corrected) | 13.979 dB |
| Return loss (raw, uncorrected) | 13.069 dB |
| VSWR (corrected) | 1.5000 |
| Expanded uncertainty U(|Gamma|), k=2 | +/- 0.01498 |
| -> return loss | 13.352 to 14.655 dB |
| -> VSWR | 1.4541 to 1.5477 |
| Mismatch uncertainty (through) | +0.0866 / -0.0875 dB |
| Standards model | ideal (Open=+1, Short=-1, Load=0) |
| OSL solve residual (numerical) | 2.22e-16 |
Computed by RF Cal Lab - VNA Error Correction + Uncertainty v1.3.0, the same plugin code the GDBS app runs, when this page was built. To run it on your own figures, open it in GDBS Pro.
Accredited RF and microwave calibration labs need OSL error correction and a defensible measurement-uncertainty budget that an assessor can reproduce, yet much of that math lives in spreadsheets or instrument firmware no one can audit line by line.
This plugin computes the textbook 1-port 3-term error correction as an exact complex solve and builds a GUM expanded-uncertainty budget entirely in the browser, with every term traceable to a published reference and no measurement data sent to any cloud.
Leave your email and we will tell you when we add or update RF & Metrology tools like this one, at most once a month.
We email you once to confirm. After that, unsubscribe from any issue in one click. Your address goes to VaultSync only (privacy).
Correct a measurement and generate the uncertainty budget client-side, then route the result into a Sandbox workspace, a Worklog case, or a Gate client portal. Nothing is uploaded to anyone's cloud.